Optimization of a Vibrating Sample Magnetometer for a laboratory physics course

Authors

  • Luis Avilés-Felix Bariloche Atomic Center - Balseiro Institute. Rio Negro, Argentina
  • Elmer Nahuel Monteblanco Départament P2M Nanomagnéstime et Electronique de Spin, Institut Jean Lamour, Université de Lorraine, Boulevard des Aiguillettes, Villerslés-Nacy, Nacy, France
  • Abel Gutarra Laboratory of Nanostructured Materials, National University of Engineering. Lima Peru

DOI:

https://doi.org/10.21754/tecnia.v26i2.55

Keywords:

Vibrating sample magnetometer, Magnetometry, Instrumentation

Abstract

This paper describes the implementation and a detailed optimization of a Vibrating Sample Magnetometer (VSM) for an undergraduate physics course laboratory. The VSM operation parameters were extensively discussed using Foner and Mallison coils configuration. The influence of the involved parameters (e.g. oscillation frequency, oscillation amplitude, rate change of the external magnetic field, coils configuration, etc.) on the induced voltage in the pick-up coils were discussed. A disk of nickel of 6-mm diameter was used for the calibration of the magnetometer, comparing the hysteresis loop measured with our magnetometer with the one obtained using a commercial VSM. Magnetization curves of two different samples were obtained in order to test the sensitivity of the magnetometer. The vibrating sample magnetometer implemented in the present work is able to detect changes in the total magnetic moment down to 10-3 emu. The detailed optimization of the VSM described in the present work is an example of how to solve a real problem in condensed matter, related to the determination of the magnetization value of a magnetic sample.

Downloads

Download data is not yet available.

References

[1] Foner S. Versatile and Sensetive Vibrating-Sample Magnetometer. Rev. Sci. Instrum. 1959; 30(7): 548-557

[2] Zieba A, Foner S. Detection coil, sensitivity function, and sample geometry effects for vibrating sample magnetometers. Rev. Sci. Instrum. 1982; 53(9): 1344-1354.

} [3] Burgei W, Pechan MJ, Jaeger H. A simple vibrating sample magnetometer for use in a materials physics course. Am. J. Phys. 2003;71(8): 825-828

Shintaku K, Daitoh Y, Shinjo T. Magnetoresistance affect and interflayer exchange coupling in epitaxial Fe/Au(100) and Fe/Au(111) multiflayers. Phys. Rev. B. 1993;47(21): 14584-14587

[5] Gredig T, Krinorotov IN, Eames P, et al. Unidirectional coercivity enhancement in ec¿xchange-biased Co/coO. Appl. Phys. Lett.2002; 81(7): 1270-1272.

Vega-Chacón J, Picasso G, Avilés-Félix L, et al. Influence of synthesis experimental parameters on the formation of magnetite nanoparticles prepared by polyol method. Adv. Nat. Sci. Nanosci. Nanotechnol. 2016;7(1): 015014.

[7] DeVore S, Gauthier A, Levy J, et al. Improving student Understanding of lock-in amplifiers. Am. J. Phys. 2016; 84(1):52-56

[8] Mallinson J. Magnetometer Coils and Reciproty. J. Appl. Phys. 1966;37(6):2514-2515

[9] Crangle J, Goodman GM. T he Magnetization of Pure Iron and Nickel.Proc.Roy.Soc. Lond. 1971; 321(1547): 477-491

Published

2016-12-01

How to Cite

[1]
L. Avilés-Felix, E. N. Monteblanco, and A. Gutarra, “Optimization of a Vibrating Sample Magnetometer for a laboratory physics course”, TECNIA, vol. 26, no. 2, p. 27, Dec. 2016.

Issue

Section

Articles