Implementation of the barium analytical technique in wastewater by atomic absorption spectrometry.

Authors

  • José W. Camero Jiménez Universidad Nacional de Ingeniería de Lima – Perú
  • Víctor A. Huamaní León EQUAS S.A.

DOI:

https://doi.org/10.21754/tecnia.v24i1.28

Keywords:

ISO/IEC 17025, atomic absorption spectrometer, barium, implementation, method detection limit, wastewater

Abstract

The implementation of a test method ensures the analytical reliability of the test result, reason why the present paper deals with the implementation of the analytical method described in the Standard Methods 3111 D - Direct Nitrous Oxide - Acetylene Flame Method for the determination of barium in the matrix of wastewater, using an atomic absorption spectrometer. It is implementation has the approach of the system of management of quality of testing laboratories ISO/IEC 17025: 2005, being evaluated statistical parameters: accuracy (trueness and precision), method detection limit, quantification limit, working range, linearity, sensitivity and robustness. 

Downloads

Download data is not yet available.

References

[1] ISO/IEC 17025:2006“General requirementsfor the competence of testing and calibrationlaboratories”.

[2] Gonzálezz, A. G.,Herrador, M. Á.,“Apractical guide to analytical methodvalidation, including measurementuncertainty and accuracy profiles”.Trends inAnalytical Chemistry, Vol. 26, No. 3, 2007.

[3] INDECOPI-CRT,“Guía para la validaciónde métodos de ensayo y las directrices para laimplementación y evaluación de métodos deensayos sensoriales”. Lima–Perú,2003.

[4] COVENIN-ISO 3534-1:1995,“Estadística.Vocabulario y símbolos. Parte 1: Términos relativos a probabilidades y estadísticageneral”.

[5] NTP-ISO 5725-1:1999, Exactitud (Veracidady Precisión) de los métodos de medición-Parte 1: Principios y definicionesgenerales. 1ªEdición.

[6] Mendenhall, W.,Scheaffer,R.l. Wackerly,D.D.,“Estadistica matemática conaplicaciones”.Grupo editorial Iberoamericana

[7] S. Kuttatharmmakul, D.L, Massart, J,Smeyers-Verbeke,“Chemometrics andIntelligent Laboratory Systems”, 52(2000)63-73

[8] Maroto Sánchez, A.,“Incertidumbre en métodos analíticos de rutina”, Universitat Rovira i Virgili, Tarragona 2002.

[9] EURACHEM,“theFitness for Purpose ofAnalytical Methods”, 1998.

[10] APHA, AWWA, WEF.“Standard Methodsfor the Examination of WaterandWastewater”, January 2012, 22 St Ed.Washington.

[11] FAO,“Guidelines for Quality Management inSoil and Plant Laboratories”. Rome, 1998,chapter 6.

[12] Miller, J.N.,Miller, J.C.,“Statistics andChemometrics for AnalyticalChemistry”.England, Pearson Education, 5editions, 2005,pp 145.

[13] AOAC,“Assoc Off Anal Chem, 1993; Huber,1999.En: AEFI, Validación de métodosanalíticos”, Cataluña, 2001, pp. 84.

[14] ISO 5725-2:1994“Accuracy (trueness andprecision) of measurement methodsandresults--Part 2: Basic method for thedetermination of repeatability andreproducibility of a standard measurementmethod”.

[15] http://www.hse.gov.uk/biocides/copr/pdfs/validation.pdf

[16] http://franklin.chem.colostate.edu/diverdi/C431/experiments/atomic_absorption_spectroscopy/Agilent%20-%20analytic%20methods.pdf

Published

2014-06-01

How to Cite

[1]
J. W. Camero Jiménez and V. A. Huamaní León, “Implementation of the barium analytical technique in wastewater by atomic absorption spectrometry”., TEC, vol. 24, no. 1, p. 21, Jun. 2014.

Issue

Section

Articles